4 Isolated power panels Complete solutions for critical care areas Isolated power panels Standard features: „„ Single-phase, low-leakage isolation transformer, with primary and secondary voltages configured at factory and rated to system requirements „„ Primary circuit breaker „„ Up to 16 branch circuit breakers (maximum allowed by UL) „„ LIM2010 line isolation monitor, featuring self-calibration, a wide variety of alarms, including total hazard current (THC, configurable for 2 mA or 5 mA), voltage, overload, and more „„ Reference ground bus Integratable options: „„ Onboard ground fault location system „„ Transformer load and temperature monitoring Standards and requirements: „„ Listed to UL 1047 (standard for isolated power systems equipment) „„ Meets requirements for NFPA 99 / CSA Z32 for isolated power systems „„ LIM2010 listed to UL 1022 (standard for line isolation moni- tors) Supported accessories: „„ MK2000 series remote indicators „„ GPM series power / ground modules „„ EDS series ground fault location equipment „„ Bender remote communication system „„ Connection to Bender remote stations, including the MK2430 and MK800